Molecular Spectroscopy

Physical Chemistry · Advanced Placement

Molecular Spectroscopy: the big picture

Spectroscopy measures transitions between quantized states, and the intensity of a transition depends on the transition dipole moment, which vanishes unless symmetry permits. The Beer–Lambert law relates absorbance to concentration, path length and molar absorption coefficient, providing the basis for quantitative analysis; deviations occur at high concentration where interactions change the absorbing species. Rotational spectroscopy in the microwave region requires a permanent dipole and yields bond lengths with great precision, with line intensities reflecting the thermal population of rotational levels and their degeneracies. Vibrational spectroscopy in the infrared requires a vibration that changes the dipole moment, and the rotational structure superimposed on a vibrational band produces branches whose spacing gives the rotational constant in both vibrational states. Raman spectroscopy depends instead on a change in polarizability and so accesses modes forbidden in the infrared, which is why the two techniques are complementary. Electronic spectroscopy involves transitions between electronic states, with vibrational structure arising from the Franck–Condon principle: the transition occurs vertically because nuclei cannot move during the electronic rearrangement, so intensity is distributed among vibrational levels according to the overlap of the vibrational wavefunctions. Excited states decay by fluorescence, intersystem crossing and phosphorescence, and these pathways determine photochemical behaviour.

The Beer–Lambert Law & Quantitative Absorption

Absorbance rises in exact proportion to concentration and path length, which is what makes ultraviolet and visible spectroscopy the workhorse of quantitative analysis. The constant of proportionality, the molar absorption coefficient, is a property of the species and the wavelength, and it can span six orders of magnitude depending on whether the transition is symmetry-allowed. The law holds only while the absorbing species is unchanged by its surroundings, so deviations at high concentration are not experimental failures but evidence that association or ionisation is occurring.

{{illustration:pchem-l08-f1}}

{{illustration:pchem-l08-f2}}

Key relations

- The Beer–Lambert law. A = ε[J]L ; A = -log₁₀I/(I₀) ; I = I₀10^(-ε[J]L)

Electronic Spectra, Franck–Condon & Excited-State Decay

An electronic transition happens so fast that the nuclei have no time to move, so the molecule arrives in the upper state at the lower state's bond length. That vertical transition is the Franck–Condon principle, and it explains why an electronic band is not a single line but a progression whose intensities follow the overlap of the vibrational wavefunctions. Once excited, a molecule can fluoresce promptly, cross to a triplet state and phosphoresce slowly, or lose the energy as heat. Which route wins decides whether a dye is a bright label, a photosensitiser, or merely a way of warming the solvent.

{{illustration:pchem-l08-f3}}

{{illustration:pchem-l08-f4}}

Key relations

- Franck–Condon factors and the Stokes shift. I ∝ |∫ ψ_(v′)*ψᵥ dτ|² ; Δν̃_Stokes = ν̃_abs − ν̃ₑₘ